7.3 Menu "Access alarm table"
In this menu, up to 20 of the last alarm
messages are listed in a table. In case of
inquiries, please provide the service
technician with this information.
The enclosed error list contains
further information.
7.4 Menu "Diagnostic programs"
Vacuum and system test
The program checks the final value (in
mbar) that the vacuum reaches after a
defined period of time and how much
time passes until 50 mbar are reached.
Muffle test
By measuring the maximum power
consumption of the heating muffle, the
'age' of the heating muffle is determined
and presented by a bar diagram. If the
muffle has a high power consumption
(good, new), the bar is solid. At a certain
stage, the user is informed that the
heating muffle has to be replaced shortly.
Keypad test
If the user conducts this test, all the keys
are represented on the display by a
matching symbol. Pressing the
corresponding key results in the symbol
of the display being inverted to change
to non-inverted (depending on the actual
condition.
F1 is not represented, since this key
is used to quit the menu.
Display test
The test merely presents a certain pattern
(e.g. chessboard pattern: inverted /
non-inverted) so that the user may check
for defective pixels.
Press drive test
Automatic check of the press drive.
Possible malfunctions are pointed out.
7.5 Temperature calibration
The accuracy of the thermocouple changes
during use. In order for it to function
optimally, it needs to be recalibrated on a
regular basis. The automatic furnace
calibration set (Automatic Temperature
Checking Set 1) has been especially
developed for this purpose. We recommend
that you check your furnace with the
Automatic Temperature Checking Set 1 after
approximately 50 press cycles or after 3 to 4
months of use. The corresponding message
will be displayed. Calibration should be
conducted while the furnace is at operating
temperature (stand-by temperature). It takes
approx. 2 hours. For that reason, we
recommend that you calibrate the furnace
overnight in order to save time.
Important information
The automatic request for
calibration after 50 press cycles is
only displayed in the "Press
furnace" mode. If the furnace is in the
"Ceramic furnace" mode, the message will
not appear after 50 firing cycles. However,
we recommend conducting a calibration
after approx. 3-4 months if the furnace is in
frequent use.
1. Insert the melting sample (122) in the
ceramic base block (121).
2. Screw the sample (122) to the contact
pins (123).
Important
Do not use tongs. Tighten until
barely fingertight.
3. Go to "Miscellaneous" and select
calibration program. Now the furnace
head opens.
4. Remove the firing plate (50) from the
furnace with the firing tongs and place it
on the sagger tray.
5. Use tongs to insert the completed test
assembly into the recesses for the ATK 1
(124).
6. Place the tongs at the center of the test
assembly and press lightly until it
perceptibly clicks into place.
7. Start the calibration program.
8. At the end of the program, remove the
test assembly from the furnace using the
tongs and allow it to cool.
Contraindication
Under no circumstances should you
pull on the test sample. The ceramic
base will break as a consequence.
9. Replace the firing plate (50) and select a
firing program. The furnace head will
close automatically.
10. Once the sample assembly has cooled,
take it apart.
11. Use new melting samples for the next
calibration procedure and begin with
step 1.
123
123
121
122
50
74