Operating InstructionsEP 600 CombiPress Furnace and Ceramic FurnaceValid as of Software Version V4.0
102.1.4Risk of crushing / burn hazardNever reach under the furnacehead during operation. There isa risk of crushing and a burnhazard.2.1.6Contraindica
112.1.9ContraindicationPress cycles must not beconducted without the firingplate (50) in place.2.1.10Contraindication The firing plate (50) must becorre
122.1.15ContraindicationThe furnace head must not becarried by holding the cables, asthere is a risk of damaging thecables and the correspondingconnec
132.2 Health and safety instructionsThis furnace has been designedaccording to EN 61010-1 andhas been shipped from themanufacturer in safe operatingco
143. Product Description3.1 ComponentsThe EP600 Combi Press andCeramic Furnace comprises thefollowing components:– Furnace base with electroniccontrol
154.1 Unpacking and checking the contentsThe new packaging provides thefollowing advantages:– Reusable packaging– Closing mechanism with inte-grated t
16The furnace components areassembled as follows:Step 1:Assembling the tray and thefiring plate– Mount the supports for thesagger tray (25)– The sagger
17Step 6:Setting-up the cooling grid forinvestment ringsThe cooling grid for investmentrings is supplied fully assembled.Please position the grid in a
18Step 7:Connecting the card readerConnect the adapter (165) withthe RS232 PC/Printer interface(14) and insert the plug (161) orthe card reader (Type
195.1 Starting the furnace5.1.1 Start-upAfter the furnace has been switched on, theunit undergoes a self-test, and if necessary, apreheating cycle. Th
2
Please use only the originalstandard programs, which areespecially coordinated with thecorresponding materials for theIvoclar Vivadent all-ceramic sys
21Internal programsWith this menu item, you may access 50 individual firing programs that are stored in the furnace.Program memory cardThis option also
22Important noteVacuum:If the value 0 is entered for V1 and V2, no vacuum will be created.Long-term coolingValues between 50 and 1200 °C (122 and2192
235.4 Introduction to the menu "Miscellaneous"This menu is called up via the main menu. It permits the selection of a number oflower-level
24Direct –>Press/Firing programs areprotocolled (Press/ Firing programprotocol table) and are shown or printedimmediately on the serial interfaceRS
255.5 Operating the menu / key functions5.5.1 Navigation1231 Information lineThis line always shows the most currenttemperature and operating mode. A
266.1 Switching on/offPut the I/O switch at the rear of the furnaceon position "I". The furnace is now switchedon and the EP600 logo appears
27What Part WhenCheck all plug-in connections for correct fit Various external connections weeklyCheck if the furnace head opens smoothly and without
287.3 Menu "Access alarm table"In this menu, up to 20 of the last alarmmessages are listed in a table. In case ofinquiries, please provide t
297.6 Changing the press plungerIn order to facilitate the changing of thepress plunger, we recommend lowering thepress plunger manually by using thec
3Table of ContentsPageViews of the Furnace, List of Parts 41 Introduction / Signs and Symbols 81.1 Preface1.2 Signs and Symbols1.3 Notes regarding the
30This chapter will help you to recognizemalfunctions and take the appropriatemeasures or, if possible, to perform minorrepairs.8.1 Error messages and
318.2 Technical malfunctionsThese malfunctions may occur without an error message being displayed:Description Double-Check ActionVacuum is not or only
32Description Double-Check ActionFinal vacuum is not reached Is the vacuum hose OK? Check vacuum hose (13) and hose connection.Was the absolute value
339.2 Technical dataPower supply: 200–240 Volt, 50–60 Hz110–120 Volt, 50–60 HzInstallation category II Max. power consumption: 12 A at 110–120 VAC8,5
349.3 Acceptable operating conditions– Acceptable temperaturerange:+5 °C to +40 °C(+41 °F to +104 °F)– Acceptable humidity rangeMaximum relative humid
3510. MiscellaneousThe program parameters are set in thefactory. The parameters used may differ,but they are based on the latest standardof technology
3610.2 Firing table4)These values are related to the possible set values. The actually achieved values however may differ and depend on various factor
Description:F1... skip into the next menu below with F1(F3)... skip back with F33710.3 Menu structure of the EP600 CombiPreheatingStartF2...F3Main men
3810.4 1-step firing program
3910.5 2-step firing program
4List of PartsA = Furnace Base 1 Sagger tray BP12 Screws for sagger tray3 Rubber foot4 Air vents5 On/Off switch6 Power socket7 Vacuum pump socket8 Ext
Ivoclar Vivadent AGBendererstrasse 2FL-9494 SchaanLiechtensteinTel. +423 235 35 35Fax +423 235 33 60www.ivoclarvivadent.comIvoclar Vivadent Pty. Ltd.1
5455441179515473123535452141020127689
656707172696855151924507480736365646267661141081041021031011001051061071091101111121135857566061628159
712552171379115475181617237322121122140123767778DE164163161160170150162165GFH
81.1 PrefaceDear Customer,Thank you for having purchasedthe EP600 Combi Press andCeramic Furnace. This furnacewith the intelligent press drivehas been
2. Safety FirstThis chapter is especially important for individualswho work with the EP600 Combi or who have tocarry out maintenance or repair work. T
Ok