Glass-ceramic Fusion Process
Remove the IPS e.max CAD Crystall./Connect with the IPS Spatula from the capsule and distribute it on the inner aspect of the IPS e.max CAD veneering structure and ...
... evenly distribute the IPS e.max CAD Crystall./Connect material by means of the vibrator.Apply IPS e.max CAD Crystall./Connect on the occlusal aspect of the IPS e.max ZirCAD frame-
work and ...
Fusion Process
The fusion process must be completed quickly in order to prevent a permature drying of the IPS e.max CAD
Crystall./Connect fusion glass-ceramic. The amount of IPS e.max CAD Crystall./Connect contained in one capsule  is 
sufficient for a  4-unit bridge. Please observe the following procedure for the fusion process:
–  Place the open capsule on the finger clip and put on  the finger clip.
–  Place some IPS e.max CAD Crystall./Connect on the occlusal aspect of  the IPS e.max ZirCAD framework in order to avoid 
hollow spaces in the fusion area. Evenly distribute the IPS e.max CAD Crystall./Connect material by vibrating it for a 
short time with the Ivomix. 
–  Remove small amounts of IPS e.max CAD Crystall./Connect with the IPS  Spatula and distribute it on all inner aspects of 
the IPS e.max CAD veneering structure.
–  Hold occlusal surface of the IPS e.max CAD veneering structure against the vibrating plate of Ivomix for a short time so 
that the IPS e.max CAD Crystall./Connect  material is evenly distributed.
–  Insert the IPS e.max ZirCAD framework in the correct position into the IPS e.max CAD veneering structure. 
–  Hold the occlusal surface of the restoration against the vibrating plate of the Ivomix. At the same time, apply slight 
pressure against the IPS e.max ZirCAD  framework, e.g. by means of the IPS Spatula, in order to evenly insert it  into the 
IPS e.max CAD veneering structure.
–  When joining the IPS e.max ZirCAD framework and the IPS e.max CAD veneering structure, the IPS e.max
CAD Crystall./Connect material must be evenly squeezed out of the entire circular fusion joint. If no fusion
glass-ceramic is being squeezed out in some places, an insufficient amount of material has been applied to
the IPS e.max CAD veneering structure, and the entire procedure must be repeated with a new capsule.
Separate and clean the IPS e.max ZirCAD framework and the IPS e.max CAD veneering structure under run-
ning water before repeating the procedure.
–  The IPS e.max CAD veneering structure is in the correct position when it sits flush on the collar of the IPS e.max  ZirCAD 
framework.
–  Important: Hold the restoration against the vibrating plate of the Ivomix  only until the IPS e.max ZirCAD framework 
and the IPS e.max CAD veneering  structure are in the correct position to one  another. A longer vibrating process may 
result in too  much IPS e.max CAD Crystall./Connect material escaping.
–  Switch off the Ivomix.
–  The IPS e.max ZirCAD framework and the IPS e.max CAD veneering  structure are in a firm position when the restoration 
is no longer being vibrated. 
Preparation Fusion process
Cleaning,
checking
Fusion/
Crystallization
firing