Pretreatment
– Removegrosscontaminationimmediatelyafteruseor
within2hours.Forthispurpose,thoroughlyrinsethe
lightprobeunderrunningwater(foratleast
10seconds).Alternatively,useasuitablealdehyde-free
disinfectingsolutiontopreventthefixationofblood.
–Toremovecontaminationmanually,useasoftbrushor
softcloth.Partiallypolymerizedcompositecanbe
removedwithalcoholandaplasticspatula,ifnecessary.
Donotusesharporpointedobjects,astheymay
scratchthesurface.
Cleaning and disinfection
Forcleaning,immersethelightprobeinacleaningsolu-
tionandensurethatitissufficientlycoveredwithliquid
(ultrasoundorcarefulbrushingwithasoftbrushmay
supporttheeffect).Aneutral-enzymaticcleaningagentis
recommended.Whencleaninganddisinfecting,please
makesurethattheagentsusedarefreeof:
–organic,mineralandoxidizingacids(theminimum
admissiblepHvalueis5.5)
–alkalinesolution(themaximumadmissiblepHvalue
is8.5)
–oxidizingagent(e.g.hydrogenperoxide)
Afterwards,removethelightprobefromthesolutionand
thoroughlyrinseitunderrunningwater(foratleast
10seconds).Cleaninginathermaldisinfectorisaneffec-
tivealternative.
Sterilization
Thoroughcleaninganddisinfectingisimperativeto
ensurethatthesubsequentsterilizationiseffective.Use
onlyautoclavesterilizationforthispurpose.Thesteriliza-
tiontime(exposuretimeatsterilizationtemperature)is
4minutesat134°C(273°F);pressureshouldbe2bar
(29psi).Drythesterilizedlightprobeusingeitherthe
specialdryingprogramofyoursteamautoclaveorhot
air.Thelightprobehasbeentestedforupto200sterili-
zationcycles.
Afterthat,checkthelightprobefordamage.Holdit
againstlight.Ifindividualsegmentsappearblack,glass
fibresarebroken.Ifthisisthecase,replacethelight
probewithanewone.
Battery contacts
Toensurereliableconductivityatalltimes,keepthe
batterycontactsfreefrompossiblecontamination(e.g.
compositeresidue).Forthispurpose,cleantheaffected
contactsregularlyinthecourseoftheusualwipe
disinfection(aftereachpatient).
Disposal
Thecuringlightmustnotbedisposedofas
urbanwaste.Disposeunserviceablebatteries
andpolymerizationlightsaccordingtothe
correspondinglegalrequirementsinyour
country.Batteriesmustnotbeincinerated.