Ivoclar-vivadent Bluephase Style User Manual Page 87

  • Download
  • Add to my manuals
  • Print
  • Page
    / 88
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 86
87
Immunity test
ConductedRF
IEC61000-4-6
RadiatedRF
IEC61000-4-3
IEC 60601 test level
3V
rms
150kHzto80MHz
3V/m
80MHzto2.5GHz
Compliance level
10 V
10 V/m
Electromagnetic environment - guidance
PortableandmobileRFcommunicationsequipmentshouldnotbeusedclosertoany
partoftheBluephaseStyle,includingcables,thantherecommendedseparationdistance
caculatedfromtheequationapplicabletothefrequencyofthetransmitter.
Recommended separation distance
d=0.35
P
d=0.35P80MHzto800MHz
d=0.70
P800MHzto2.5GHz
wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)according
tothetransmittermanufactureranddistherecommendedseparationdistanceinmeters
(m).
FieldstrengthsfromfixedRFtransmitters,asdeterminedbyanelectromagneticsitesur
-
vey,
a
shouldbelessthanthecompliancelevelineachfrequencyrange.
b
Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowingsymbol:
Guidance and manufacturer‘s declaration - electromagnetic immunity
TheBluephaseStyleisintendedforuseintheelectromagneticenvironmentspecifiedbelow.ThecustomerortheuseroftheBluephaseStyleshould
assurethatitisusedinsuchanenvironment.
NOTE1:At80MHzand800MHz,thehigherfrequencyrangeapplies.
NOTE2:Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationis
affectedbyabsorptionandreflectionfromstructures,objectsandpeople.
a
Fieldstrengthsfromfixedtransmitters,suchasbasestationsforradio(cellular/cordless)
telephonesandlandmobileradios,amateurradio,AMandFMradiobroadcastandTV
broadcastcannotbepredictedtheoreticallywithaccuracy.Toassesstheelectromagnetic
environmentduetofixedRFtransmitters,anelectromagneticsitesurveyshouldbe
considered.IfthemeasuredfieldstrengthinthelocationinwhichtheBluephaseStyle
isusedexceedstheapplicableRFcompliancelevelabove,theBluephaseStyleshould
beobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additional
measuresmaybenecessary,suchasreorientingorrelocatingtheBluephaseStyle.
b
Overthefrequencyrange150kHzto80MHz,fieldstrengthshouldbelessthan10V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the “Bluephase Style”
TheBluephaseStyleisintendedforuseintheelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.Thecustomerorthe
useroftheBluephaseStylecanhelppreventelectromagneticinterferencebymaintainingamiminumdistancebetweenportableandmobileRF
communicationsequipment(transmitters)andtheBluephaseStyleasrecommendedbelow,accordingtothemaximumoutputpowerofthe
communicationequipment.
Rated
maximum out-
put power of
transmitter
W
0.01
0.1
1
10
100
150 kHz to
80 MHz
d=0.4P
0.04
0.13
0.40
1.3
4.0
Separation distance according to
frequency of transmitter
m
80 MHz to
800 MHz
d=0.4P
0.04
0.13
0.40
1.3
4.0
800 MHz to
2.5 GHz
d=0.7
P
0.07
0.22
0.70
2.2
7.0
Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommended
separationdistancedinmetres(m)canbedeterminedusingtheequationapplicableto
thefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthe
transmitterinwatts(W)accordingtothetransmittermanufacturer.
NOTE1:At80MHzand800MHz,theseparationdistanceforthehigherfrequency
rangeapplies.
NOTE4:Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationis
affectedbyabsorptionandreflectionfromstructures,objectsandpeople.