Direct –> Press/Firing programs are
protocolled (Press/ Firing program
protocol table) and are transmitted or
printed immediately on the serial
interface RS232. Please note that
therefore, a corresponding output device
(printer or PC) has to be selected in the
menu 'Interface'.
Pager 1
Yes/No
Laboratory name
Name which will appear in the header
or footer of the various protocols.
5.3.6 Menu "Maintenance"
This menu provides the commands necessary
during the maintenance of the furnace.
Lower press plunger manually
Lift press plunger manually
5.3.7 Menu "Enter new program"
The "Main menu pressing" also provides
three individual programs (Program No. 5–7)
in addition to the standard programs. If
these programs are not shown in the menu,
they are still concealed. They can be inserted
using the following function:
To insert one of the programs (program No.
5–7), select this menu item (F4), type in the
correct code, and confirm with ENTER.
Code Program No.
14000 5
15000 6
16000 7
If the code was typed in correctly, the "Main
menu pressing" displays the new program.
By therein entering the code, a
concealed program will be released
and an already released program
will be concealed.
5.3.8 Menu "Request alarm table"
See Chapter 7.3.
5.3.9 Menu "Protocols"
After selecting this menu the table with the
calibration data is shown.
Key F4 can be used to switch to the press
program protocol table (keys to be pressed:
F4, ENTER).
In this table the protocols of the last
15 press programs are stored. These
protocols can be transmitted manually from
here to the PC or printer (this depends on
the settings in the menu ‘Interface').
5.3.10 Menu "Timer"
In this menu, the timer may be programmed.
With this feature, the furnace may be
heated up to stand-by temperature or
switched off at any given time.
In the interval between the switch-on and
the switch-off times, the unit is kept at
stand-by temperature.
To change the setting, select the
corresponding switch-on/switch/off time
with the cursor (F2/F3) and use the numeric
keypad for selecting the desired time.
The timer can also be switched on or off
with the key ‘timer active-yes/no’.
The times are typically typed in via the
numeric keypad. F3 is used to change the
selected menu item. ESC is used to undo
changes. ENTER is used to save the changes.
The mains switch must not be switched off
during this procedure.
When the timer is active, i.e. when
the heater is switched off, the
symbol is shown on the display. If
a program has not been completely
finished at the time when the furnace is
scheduled to turn off, the program will
nonetheless be completed before the heater
is deactivated. While the timer is active,
programs can only be started once the time
settings have been adjusted accordingly.