14
What If …
Indicator Causes Error rectification
All LEDs
orange
Thedeviceisoverheated. Allowthedevicetocooldownandtryagainaftera
certaintime.Iftheerrorpersists,pleasecontactyour
dealeroryourlocalServiceCentre.
All LEDs
red
Electroniccomponentofthehandpieceis
defective.
Removeandreinsertthebattery.Iftheerrorpersits,
pleasecontactyourdealeroryourlocalServiceCentre.
The charging
base is not
illuminated
during
charging
– Powerpacknotconnectedordefective
– Batteryfullycharged
Checkifthepowerpackiscorrectlypositionedinthe
chargingbaseorifthepowerpackisconnectedtothe
powersupplybymeansofthepowercord(LEDonthe
powerpacklightsupingreenifitworkscorrectly).
No indicator
activity of the
device with the
battery in place
Batteryempty Placethedeviceinthechargingbaseandchargefor
atleast2hours.
Batterycontactsdirty Removebatteryandcleanthebatterycontacts.